FORCE TEN
FDX FORCE GAGE
10
FORCE TEN
FDX FORCE GAGE
11
B. Keypad
B.1 Keypad Description
FDX is operated with four button keypad that controls all functions.
ON/OFF Turns FDX on and off.
Self-tests display if held down when turning gage on.
At turn-on, display ashes Force Cell capacity and
software version.
If there is no display or if low battery is indicated, the battery may be low or not
securelyconnected.ConnectingtheACPowerAdapterwillconrmthis.
ZERO Returns display to zero in non-peak and peak mode.
Returns display to zero with tare force or weight applied to the
load shaft.
PEAK Activates peak mode and recalls peak readings.
UNITS Selects units of measurement: lbf, kgf, N, or ozf.
Eachofthebuttons,exceptON/OFF,alsohasset-upfunctions.
B.2 Keypad Control
Three black keypad buttons have the following additional functions:
SCROLL Review menu options.
Review selections within each option.
SELECT Select displayed item.
Save selection.
ESCAPE Cancel and return to previous menu option.
Exit set-up menu.
Prior to actual test use, gain keypad and menu familiarity with practice.
KEYPAD MENU OPERATIONS
C. Menu Set-Up
C.1 Menu Description
FDX features options for selection in the set-up menu. At rst level are options; at
second level are selections within the options:
TABLE 4 MENU OPTIONS AND SELECTIONS
Display Option - Level One Selections - Level Two
AoFF Automatic Off No* or 30 minutes
PSS Peak Sampling Rate
Sampling rate: 100 samples per second.*
Input bandwidth: 9Hz in non-peak
33Hz in peak.
Sampling rate: 1000 samples per second.
Input bandwidth: 90Hz in non-peak
330Hz in peak.
L CAL Auto - Calibration See E.3 Auto-Calibration
* Factory default settings.
Inputbandwidth:Thebandoffrequenciesthatlteringpassesthrufromtheloadcell
to the display. All other background noise or interference frequencies are excluded.