Filtering of peak readings may be set as follows:
FP1 Average of 1 sample (No filtering)
FP2 Average of 2 samples
FP4 Average of 4 samples
FP8 Average of 8 samples (High filtering)
c) Setting the Digital Filters - Fc and FP
After arriving at "Fltd" in the menu option, press SELECT to display
the previous Fc setting (Fc 1, Fc 2, Fc 4 or Fc 8). To change the
Fc setting press SCROLL until the desired setting is displayed.
Press SELECT and "donE" will appear. DO NOT press SELECT
again to save this setting.
Proceed immediately to the FP settings by pressing SCROLL after
setting the Fc value ("donE"). The previous FP setting will appear
(FP 1, FP 2, FP 4 or FP 8). To change the FP setting, press SCROLL
until the desired setting is displayed. Press SELECT and "donE" will
appear.
Now press SELECT again to save both the Fc and FP settings.
5. Force Cell Capacity (LoAd)
The capacity of the Force Cell Module (FCM) attached to the FDV
must be entered into the setup menu (LoAd). After the “LoAd” option
is chosen the presently set value is displayed first.
If the Force Cell Module (FCM) has been changed from one capacity
to another, the new FCM capacity must be selected and saved. The
choices are 1, 2, 5, 10, 30, 50, and 100 lb capacities (see Section G:
Interchanging Force Cell Modules).
CAUTION: "LoAd" will appear on the display when the gage is first
activated if the gage has been disconnected from the battery or
AC power for an extended period of time, and the set-up menu
has reverted to default settings. User must reset "LoAd" (see