H. CALIBRATION
To ensure accuracy in force measurement, the FDV Digital Force
Gage should be tested periodically to confirm that it is within
tolerance.
H.1 Testing Calibration
The FDV Digital Force Gage is accurate to ±0.3% of the gage's full
scale capacity. To test calibration, a four point test is recommended, at
weights equal to 25%, 50%, 75%, and 100% of the gage capacity. Four
test weights, corresponding to the above percentages, are required. If
the displayed weight should differ from the value of the test weight by
more than 0.3%, the gage is out of calibration. Remember to tare any
attachments prior to testing by pressing ZERO with the FDV shaft and
hook attachment pointing down.
RECOMMENDATION: Warm-up time for the FDV is recommended.
After initial power-on, allow 2-3 minutes for the electronics to stabilize.
This will improve accuracy of reading.
Since the FDV Force Gage consists of two separate parts, the Force
Display Module (FDM) and the Force Cell Module (FCM), the total
accuracy of the gage depends on the tolerance to specifications of
either one or both of the modules. If the FDV does not meet the rated
0.3% accuracy, the gage should be returned to Wagner Instruments for
evaluation and recalibration of the modules. Call Wagner Instruments
for instructions prior to shipping.
H.2 Auto-Calibration
Factory calibration is recommended but in special situations the gage
may be calibrated in the field using the Auto-Calibration feature.
1. If the FDV is used ONLY as a non-interchangeable unit - only one
Force Cell Module is used with only one Force Display Module - then
the unit may be calibrated using the procedure outlined in Section H.3.